Eppendorf has released a software update that permanently removes VNC access from the controller. Users should download and apply this update from: https://www.eppendorf.com/software-downloads. https://www.eppendorf.com/software-downloads All affected BioFlo 320 systems always shipped with Virtual Network Computing (VNC) disabled by default, and VNC can only be enabled locally at the tower. Eppendorf has removed VNC configuration information from all current documentation, so it no longer appears in BioFlo 320 Operating Manuals. Eppendorf recommends user do the following: - Verify that VNC is disabled on the controller - Enable security so that only Admin and Supervisor roles can change VNC settings - Install Version 5.0 Software as soon as possible
Workaround
No workaround given by the vendor.
Tue, 26 May 2026 19:30:00 +0000
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
ssvc
|
Tue, 26 May 2026 18:00:00 +0000
| Type | Values Removed | Values Added |
|---|---|---|
| Description | Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted. | |
| Title | Eppendorf BioFlo 320 Use of hard-coded password | |
| Weaknesses | CWE-259 | |
| References |
| |
| Metrics |
cvssV3_1
|
Projects
Sign in to view the affected projects.
Status: PUBLISHED
Assigner: icscert
Published:
Updated: 2026-05-26T18:39:20.583Z
Reserved: 2026-04-27T18:37:39.380Z
Link: CVE-2026-7251
Updated: 2026-05-26T18:39:14.922Z
Status : Awaiting Analysis
Published: 2026-05-26T18:16:55.347
Modified: 2026-05-26T19:06:14.330
Link: CVE-2026-7251
No data.
OpenCVE Enrichment
Updated: 2026-05-26T19:45:06Z